FDA Adverse Event
Injury
Summary report: N
ATLAS II PLUS DR
MDR report key: 3910954
·
Received July 2, 2014
Report
- Report Number
- 2938836-2014-12957
- Event Type
- Injury
- Date Received
- July 2, 2014
- Date of Event
- May 13, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GB
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED FIELD EVENT OF BACKUP VVI WAS CONFIRMED IN THE LABORATORY. REVIEW OF THE DEVICE IMAGED INDICATED THAT A POWER-ON RESET HAD OCCURRED. THE DEVICE WAS TESTED USING AUTOMATED TEST EQUIPMENT, AND NO ANOMALY WAS FOUND. THE CAUSE OF THE POWER-ON RESET COULD NOT BE DETERMINED.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS FOUND TO BE IN BACK UP VVI MODE WITH NO DEFIBRILLATION CAPABILITIES. THE DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387021 | ATLAS II PLUS DR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST. JUDE MEDICAL, INC., CRMD | V-268 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |