INFUSOR
Report
- Report Number
- 1416980-2014-21369
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 11, 2014
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K071222
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER
Narratives
(B)(4). (B)(6). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF THE INVESTIGATION, OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THE DEVICE WAS MANUFACTURED FEBRUARY 8, 2014 ¿ FEBRUARY 10, 2014. EVALUATION SUMMARY: ONE SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION NOTED A WHITE THREAD-LIKE PARTICLE 4.5 MM IN LENGTH HANGING FROM THE STRESS-MEMBER FILTER. THE PARTICLE WAS IDENTIFIED TO BE A POLYESTER MATERIAL VIA FOURIER TRANSFORM INFRARED (FT-IR) SPECTROPHOTOMETER SCANNING. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE CAUSE OF THE PARTICULATE MATTER WAS DUE TO THE STORAGE AND HANDLING OF THE STRESS MEMBER SUBASSEMBLIES. THERE IS A CAPA OPEN TO FURTHER INVESTIGATE THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WAS PARTICULATE MATTER IN A LARGE VOLUME INFUSOR. THIS WAS NOTED AFTER FILLING WITH FLUOROURACIL. THE REPORTER DESCRIBED THE PARTICULATE MATTER AS A LOOSE FIBER THAT WAS ¿ATTACHED TO THE FILTER ON THE STRESS MEMBER.¿ THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387422 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 14B027 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |