ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-04618
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 25, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE TISSUE PAD DAMAGED, MELTED, AND A SMALL PORTION WAS NOT RETURNED. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND IT ACTIVATED DURING FUNCTIONAL TESTING. PROBABLE CAUSES OF TISSUE PAD DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. AVOID ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO PREVENT DAMAGE TO THE TISSUE PAD.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE WHITE TISSUE PAD FELL OFF. THE FALLEN PIECE DIDN'T FALL INTO THE PATIENT'S BODY. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 386431 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |