FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3910867 · Received July 2, 2014

Report

Report Number
1416980-2014-21354
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
December 21, 2013
Report Date
June 11, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).THE DEVICE WAS EVALUATED AND THE ISSUE WAS IDENTIFIED DURING THE REVIEW OF THE EVENT HISTORY LOGS. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE CAUSE WAS DETERMINED TO BE UNEXPECTED HIGH ULTRAFILTRATION RELATIVE TO OTHER THERAPIES. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 23:51:35. DURING NIGHT DRAIN CYCLE ONE, THE PATIENT'S ULTRAFILTRATION READING WAS 1496ML, INDICATING THE HOME PATIENT (HP) DRAINED 1496ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.