HOMECHOICE PRO
MDR report key: 3910867
·
Received July 2, 2014
Report
- Report Number
- 1416980-2014-21354
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- December 21, 2013
- Report Date
- June 11, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).THE DEVICE WAS EVALUATED AND THE ISSUE WAS IDENTIFIED DURING THE REVIEW OF THE EVENT HISTORY LOGS. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE CAUSE WAS DETERMINED TO BE UNEXPECTED HIGH ULTRAFILTRATION RELATIVE TO OTHER THERAPIES. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 23:51:35. DURING NIGHT DRAIN CYCLE ONE, THE PATIENT'S ULTRAFILTRATION READING WAS 1496ML, INDICATING THE HOME PATIENT (HP) DRAINED 1496ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.