GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00567
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 16, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION INDICATED THIS MDR MANUFACTURER REPORT NUMBER IS A DUPLICATE OF 2937094-2014-00522.
FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER EXHIBITED A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE ROTATED INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITED MODERATE DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITED MODERATE BURNT ON DETRITUS; THE OUTER FLOW TUBING EXHIBITED MILD CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS, THE POTENTIAL FOR FORWARD FIRING MAY ALSO EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER.
THREE SURGICAL FIBERS WERE REPORTED TO HAVE BEEN USED DURING A GREENLIGHT BPH PROCEDURE OF A LARGE, APPROXIMATELY 200 G, PROSTATE. AT 29,028 JOULES WHILE USING THE FIRST SURGICAL FIBER, PROSTATIC CALCIFICATIONS/STONES WERE ENCOUNTERED RESULTING IN FIBER DAMAGE AND FORWARD-FIRING OF THE OUTPUT BEAM. THE FIBER WAS EXCHANGED AND THE CASE CONTINUED USING THE SECOND FIBER, WHICH WAS REPORTED ALSO TO HAVE FAILED AT A LOW JOULE COUNT. THE PROCEDURE WAS COMPLETED USING A THIRD SURGICAL FIBER. THE PROCEDURE OUTCOME WAS REPORTED AS "OK" - NO PATIENT INJURY WAS REPORTED. THIS REPORT IS FOR THE FIRST SURGICAL FIBER USED. TIME EXPENDED: 6:27.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 386359 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 405A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |