FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3910786 · Received July 2, 2014

Report

Report Number
3005075853-2014-04612
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED, EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN11 AN ALERT SCREEN WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN ALERT SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP LOW ANTERIOR RESECTION, THE BLADE WAS BROKEN OFF SOON. THE BROKE PIECE FELL INTO THE PATIENT, BUT IT WAS RETRIEVED FROM THE PATIENT WITH A FORCEPS VIA A TROCAR. THERE WAS NO BLEEDING DUE TO THE EVENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE DOCTOR COMMENTED THAT THE DEVICE HAD NOT CONTACTED WITH A FORCEPS. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
386296 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE