FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3910764 · Received July 2, 2014

Report

Report Number
2024168-2014-04333
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
June 6, 2014
Report Date
June 12, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE SHAFT SEPARATION AND KINK WERE ABLE TO BE CONFIRMED. THE FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON A VISUAL AND DIMENSIONAL/ INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE ELECTRONIC COMPLAINT DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

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

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
387172 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 3020641

Patients

Seq Age Sex Outcome Treatment
1 GUIDE WIRE: SION