FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3910735 · Received July 2, 2014

Report

Report Number
1416980-2014-21331
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
June 10, 2014
Report Date
June 10, 2014
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. FUNCTIONAL TESTING REVEALED THAT THE DEVICE HAD PRESENTED AN LOW BATTERY ALARM. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE AN INOPERATIVE MAIN BATTERY. TO CORRECT THE CONDITION, THE MAIN BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO HAVE PRESENTED A LOW BATTERY ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
387111 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1