FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT ? LONG

MDR report key: 3910623 · Received July 2, 2014

Report

Report Number
2134265-2014-03772
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
June 9, 2014
Report Date
June 9, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: THE DEVICE WAS RETURNED FOR EVALUATION. A VISUAL EXAMINATION OF THE CRIMPED STENT FOUND STENT STRUTS ON THE PROXIMAL END OF THE CRIMPED STENT WERE DISTORTED AND LIFTED UPWARDS FROM THE STENT PROFILE. THE BUMPER TIP OF THE DEVICE SHOWED NO SIGNS OF DAMAGE. THE BALLOON CONES PROFILES WERE REVIEWED AND NO ISSUES WERE NOTED WITH THE OVERALL BALLOON PROFILE. THE BALLOON WINGS WERE TIGHTLY WRAPPED AND EVENLY FOLDED AND WERE NOT SUBJECTED TO POSITIVE PRESSURE. A VISUAL AND TACTILE EXAMINATION FOUND MULTIPLE KINKS ALONG THE HYPOTUBE SHAFT. A VISUAL AND TACTILE EXAMINATION FOUND NO ISSUES WITH THE PROFILE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/ PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT STENT DAMAGE OCCURRED. A 3.00X38MM PROMUS ELEMENT LONG DRUG-ELUTING STENT WAS SELECTED FOR USE AND ADVANCED TO TREAT THE UNSPECIFIED TARGET VESSEL. HOWEVER, IT WAS NOTICED THAT THE HEAD OF THE STENT BODY WAS LIFTED AND THE DEVICE FAILED TO CROSS THE LESION. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
387720 PROMUS ELEMENT ? LONG STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911338300 16349960

Patients

Seq Age Sex Outcome Treatment
1