V60 VENTILATOR
Report
- Report Number
- 2031642-2014-00599
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Report Date
- June 7, 2014
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- RESPIRATORY THERAPIST
Narratives
DATE RECEIVED BY MFR DATE: 08/29/2014. THIS DA BOARD WAS TESTED AND NO FAILURES WERE IDENTIFIED. THIS REPORT IS BEING SUBMITTED AS PART OF THE REMEDIATION FOR (B)(4).
THE CUSTOMER REPORTED THAT THE VENTILATOR ALARMED AND SHUTDOWN DURING NORMAL VENTILATION OPERATION DUE TO A VENT INOP DATA ACQUISITION PCBA ADC FAILURE. THE CUSTOMER REPORTED THE DEVICE WAS IN USE ON A PATIENT AND THERE WAS NO PATIENT HARM. A VENT INOP CONDITION DURING OPERATION WILL RESULT IN A VISUAL AND AUDIBLE ALARM AND PRECLUDES CONTINUED SAFE OPERATION OF THE VENTILATOR. THE USER MUST PROVIDE ALTERNATIVE VENTILATION AND HAVE THE VENTILATOR SERVICED. THE MANUFACTURERS FIELD SERVICE ENGINEER (FSE) WAS ABLE TO DUPLICATE THE REPORTED PROBLEM. THE FSE REPORTED THE DATA ACQUISITION PCB BOARD WAS REPLACED TO CORRECT THE REPORTED PROBLEM. FINAL APPLICABLE TESTING WAS COMPLETED PER OPERATING SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387197 | V60 VENTILATOR | VENTILATOR, CONTINUOUS | MNT | RESPIRONICS CALIFORNIA, INC. | V60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |