FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3910282 · Received July 2, 2014

Report

Report Number
1416980-2014-21276
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
May 27, 2014
Report Date
June 9, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 22:30:02. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 1016ML, INDICATING THE HOME PATIENT (HP) DRAINED 1016ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
386621 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 46 YR