HEARTWARE® VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2014-00682
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- January 16, 2014
- Report Date
- June 2, 2014
- Manufacturer
- HEARTWARE, INC
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1607-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
PRELIMINARY EVALUATION AND TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL. THIS FINDING WAS RE-ASSESSED PER OUR SOP REQUIREMENTS, AND DETERMINED TO BE REPORTABLE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF COMPLETION OF THE INVESTIGATION. THIS EVENT RELATES TO THE FSCA Z-1607-2014 ISSUED BY HEARTWARE, INC. FOR THIS MATTER. EVALUATION IN PROGRESS.
THE CONTROLLER AND BATTERIES WERE RETURNED; VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE CONTROLLER AND BATTERIES IN RELATION TO THE REPORTED EVENT. THOROUGH EXTERNAL VISUAL INSPECTION OF THE CONTROLLER AND BATTERIES REVEALED NO SIGNS OF PHYSICAL DAMAGE OR CONTAMINATION. THE CONTROLLER AND FOUR BATTERIES PASSED FUNCTIONAL TESTING, HOWEVER, FUNCTIONAL ANALYSIS REVEALED THAT (B)(4) CONTAINED A FAULTY CELL PAIR, WHICH MOST LIKELY CONTRIBUTED TO THE REPORTED EVENT. AN INTERNAL INVESTIGATION (CAPA) HAS BEEN OPENED TO ADDRESS THE ISSUE.
APPROXIMATELY ONE YEAR AND FOUR MONTHS POST HVAD IMPLANTATION, THE SITE REPORTED THAT THE PATIENT EXPERIENCED POWER SOURCE CHANGE IN THE CONTROLLER EARLIER THAN EXPECTED. THE BATTERY WAS REMOVED FROM THE PATIENT AND A NEW BATTERY WAS SUPPLIED. NO HARM OR INJURY TO THE PATIENT WAS REPORTED AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 386531 | HEARTWARE® VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM, BATTERY | DSQ | HEARTWARE, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4)- BATTERY| (B)(4)- BATTERY| (B)(4)- BATTERY| (B)(4)- CONTROLLER| (B)(4)- BATTERY |