FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CAUTERY INSTRUMENT

MDR report key: 3909125 · Received July 2, 2014

Report

Report Number
2955842-2014-04053
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
May 19, 2014
Report Date
June 12, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S MAIN TUBE HAD SCRATCH MARKS AND ABRASIONS. THE MAIN TUBE INSULATION WAS SCRATCHED AND SOME OF THE INSULATION WAS REMOVED AT THE DISTAL END. THE SCRATCHES MEASURED APPROXIMATELY .022 - .096 IN LENGTH. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE MAIN TUBE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DA VINCI SURGICAL PROCEDURE, IT WAS NOTED THAT THE DISTAL INSULATION ON THE MONOPOLAR CAUTERY INSTRUMENT WAS TORN. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
388134 MONOPOLAR CAUTERY INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420142-04 S10120907 275

Patients

Seq Age Sex Outcome Treatment
1 75 YR