FDA Adverse Event Malfunction Summary report: N

SYMBIQ DCHANNEL 3.01

MDR report key: 3905878 · Received May 29, 2014

Report

Report Number
9615050-2014-03714
Event Type
Malfunction
Date Received
May 29, 2014
Date of Event
January 1, 2014
Report Date
April 30, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE ALARMED WITH A S421 (MOTOR ERROR-PMC, RIGHT) MALFUNCTION ALARM CODE. THE PROBABLE CAUSE WAS A LOOSE MOTOR DUE TO THE UNDER APPLICATION OF THE ADHESIVE SEAL OR THE ADHESIVE SEAL BEING APPLIED OVER GREASE. DURING A REVIEW OF THE DEVICE HISTORY A S321 ALARM "PUMP QEP COUNT LOW" WAS NOTED. THE PROBABLE CAUSE MAY BE DUE TO A BROKEN ENCODER ON THE MOTOR. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S421 (MOTOR ERROR-PMC, RIGHT) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED "ALARMS MALFUNCTION S421". NO INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
317170 SYMBIQ DCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK