FDA Adverse Event Malfunction Summary report: N

TIP COVER ACCESSORY

MDR report key: 3905864 · Received July 1, 2014

Report

Report Number
2955842-2014-04007
Event Type
Malfunction
Date Received
July 1, 2014
Date of Event
May 30, 2014
Report Date
June 6, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K112263
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE TIP COVER ACCESSORY WAS RETURNED IN THE ORIGINAL SEALED POUCH. MOST LIKELY, THE WRONG ITEM WAS SENT BACK. THE INSTRUMENT ACCESSORY IS DESIGNED AS A SINGLE USE ACCESSORY AND WAS FOUND TO BE UNDAMAGED. WARNING: FAILURE TO INSTALL THE TIP COVER ACCESSORY PROPERLY MAY RESULT IN: IMPROPER SCISSOR OPENING; TIP COVER ACCESSORY FALLING OFF; ELECTRICAL ARCS AND ALTERNATE SITE BURNS THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WITH THE MCS TIP COVER ACCESSORY INSTALLED DURING A DA VINCI HYSTERECTOMY PROCEDURE, THE TIP COVER STARTED SLIDING OFF OF THE INSTRUMENT. ACCORDING TO THE CUSTOMER, THE TIP COVER CAN BE TAKEN OFF EASILY BY HAND WITHOUT USE OF THE REMOVAL TOOL. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
382744 TIP COVER ACCESSORY ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 400180-12 M10140408

Patients

Seq Age Sex Outcome Treatment
1