FENESTRATED BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04008
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- June 2, 2014
- Report Date
- June 6, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE PITCH CABLE WAS BROKEN AT THE PROXIMAL CLEVIS HUB. THE CABLE SEGMENT STUCK OUT AT THE INSTRUMENT'S WRIST. NO OTHER DAMAGE OR WEAR MARKS WERE OBSERVED ON THE CLEVIS. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY PROCEDURE, A WIRE ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS FRAYED ON ONE SIDE AND LOOPED ON THE OTHER. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 382743 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M11140206 440 |