FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR

MDR report key: 3905843 · Received July 1, 2014

Report

Report Number
3015876-2014-00747
Event Type
Malfunction
Date Received
July 1, 2014
Date of Event
June 3, 2014
Report Date
June 3, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K103567
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED MAIN KEYPAD ASSEMBLY AND DETERMINED THAT THE RIGHT SIDE OF THE CHARGE KEY WAS DAMAGED AND NEARLY SHORTED. THIS WOULD CAUSE THE CHARGE KEY TO INTERMITTENTLY ACTIVATE WHEN THE ENERGY UP/DOWN OR SHOCK BUTTONS WERE PUSHED, WHICH PROHIBITED THE FUNCTION OF THESE KEYS.

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO OBSERVED THAT THE SHOCK KEY WAS REGISTERING AS BEING PRESSED ALL THE TIME, CAUSING THE CONNECT ELECTRODES MESSAGE. PHYSIO THEN REPLACED THE MAIN KEYPAD ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR DEVICE WAS INTERMITTENTLY DISPLAYING A "CONNECT ELECTRODES" MESSAGE. WHEN THE DEVICE IS DISPLAYING THIS MESSAGE, THE DEVICE COULD POTENTIALLY NOT HAVE THE ABILITY TO DELIVER DEFIBRILLATION THERAPY TO A PATIENT. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE. UPON EVALUATION OF THE DEVICE, IT WAS OBSERVED THAT THE SHOCK BUTTON WAS ALWAYS REGISTERING AS BEING PRESSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
382742 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC 15

Patients

Seq Age Sex Outcome Treatment
1