FDA Adverse Event Injury Summary report: N

TELIGEN

MDR report key: 3905826 · Received July 1, 2014

Report

Report Number
2124215-2014-12770
Event Type
Injury
Date Received
July 1, 2014
Date of Event
May 15, 2014
Report Date
July 29, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Removal / Correction Number
Z-0026-2014
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY UNDERGOING ANALYSIS. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH THE DEVICE¿S ABILITY TO DELIVER THERAPY WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. REVIEW OF THE DEVICE MEMORY INDICATED THAT A VOLTAGE ALERT WAS RECORDED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE¿S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL. ON AUGUST 29, 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS A PART OF THE IDENTIFIED POPULATION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI). IT WAS THOUGHT THAT THE DEVICE MAY HAVE EXPERIENCED PREMATURE BATTERY DEPLETION. A SAVE TO DISK WAS SENT IN TO BOSTON SCIENTIFIC TECHNICAL SERVICES FOR ANALYSIS. ANALYSIS OF THE DATA DISK REVEALED THAT THE DEVICE WAS MALFUNCTIONING. THE PATIENT WAS SEEN FOR AN EXPLANT PROCEDURE WHERE THE DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE DEVICE HAS BEEN RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
382546 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization| L| R 4087| 0157| T125| E110