FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SINGLE CHANNEL
MDR report key: 3905812
·
Received May 29, 2014
Report
- Report Number
- 9615050-2014-03722
- Event Type
- Malfunction
- Date Received
- May 29, 2014
- Date of Event
- April 28, 2014
- Report Date
- May 1, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE PROXIMAL OCCLUSION TEST. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS AVAILABLE FOR CLINICAL USE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 317500 | SYMBIQ SINGLE CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |