SCREWDRIVER DUO, SW 3.5 QUICK COUPLING
MDR report key: 3905788
·
Received May 28, 2014
Report
- Report Number
- 8043862-2014-00006
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- April 30, 2014
- Report Date
- May 27, 2014
- Manufacturer
- AAP IMPLANTATE AG
- Product Code
- HXX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE SIXHEAD OF THE SCREWDRIVER BROKE OF DURING INSERTION OF THE PLATE. THE PART WHICH BROKE OFF HAS BEEN NOT RETURNED TO THE MFR. THE SCREWDRIVER IS ABOUT TWO YEARS OLD. DEPENDING ON THE QUANTITY OF USAGES AND THE FORCES THAT APPLY A WEAR OUT COULD OCCUR AND A BREAKAGE IS THE RESULT. THE SCREWDRIVER SHOWED NO FURTHER DAMAGES.
Description of Event or Problem · 1
DRILL BIT BROKE OFF DURING SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314836 | SCREWDRIVER DUO, SW 3.5 QUICK COUPLING | SCREWDRIVER | HXX | AAP IMPLANTATE AG | IU 7835-65 | I004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |