TWIST DRILL D3.8, L 310, COIL 50, QUICK COUPLI
MDR report key: 3905768
·
Received May 28, 2014
Report
- Report Number
- 8043862-2014-00004
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- April 30, 2014
- Report Date
- May 27, 2014
- Manufacturer
- AAP IMPLANTATE AG
- Product Code
- HTW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DRILL HAS BEEN RETURNED AND THE TOP HAS BEEN BROKEN OFF. THE TOP OF THE DRILL WAS NOT RETURNED. THE SHAFT OF THE DRILL SHOWS SEVERAL DAMAGES. PROBABLY A FOREIGN BODY IN THE DRILL GUIDE CAUSED THIS DAMAGES WHICH RESULTED IN THE BREAKAGE OF THE DRILL.
Description of Event or Problem · 1
DRILL BIT BROKE OFF DURING SURGERY. IT IS NOT KNOWN IF THE TOP OF THE DRILL WHICH BROKE OFF REMAINS IN THE BODY.