ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2014-11635
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- May 18, 2014
- Report Date
- June 27, 2014
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
BOSTON SCIENTIFIC RECEIVED ADDITIONAL INFORMATION THAT THE DEVICE AND LEAD WERE LATER EXPLANTED AND REPLACED. IT WAS ALSO NOTED THAT THIS LEAD HAD EXHIBITED INCREASED PACING THRESHOLD MEASUREMENTS AT THE TIME OF THE DEVICE REPLACEMENT PROCEDURE; THEREFORE, AT THAT TIME, THE RATE/SENSE PORTION OF THE LEAD WAS SURGICALLY ABANDONED AND A NEW RV PACING LEAD WAS IMPLANTED. THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE LEAD WAS NOTED TO BE SEVERED 140MM FROM THE TERMINAL PIN. TWO SEGMENTS WERE RETURNED THAT MEASURED 630MM IN TOTAL, INDICATING A 10MM PORTION OF THE LEAD WAS MISSING. AN EXTRACTING STYLET WAS INSERTED INTO THE TIP SEGMENT, AND SIGNIFICANT EXPLANT DAMAGE WAS OBSERVED ON BOTH LEAD SEGMENTS, INCLUDING STRETCHED CONDUCTOR COILS AND TORN AND CUT INSULATION. TISSUE AND CALCIFICATION WERE ADHERED TO THE PROXIMAL AND DISTAL SHOCKING COILS, AND THE CALCIFICATION WAS ALSO NOTED ON THE HELIX. THE HELIX WAS EXTENDED. ELECTRICAL CONTINUITY TESTING WAS PERFORMED ON THE RETURNED LEAD SEGMENTS, AND ALL PATHWAYS WERE CONFIRMED TO BE ELECTRICALLY CONTINUOUS. NO CONDUCTOR FRACTURES WERE OBSERVED. LABORATORY ANALYSIS CONCLUDED THAT THE BUILD-UP OF CALCIFICATION ON THE SHOCKING COILS HAD CREATED A NON-CONDUCTIVE BARRIER, CAUSING THE SHOCK IMPEDANCE MEASUREMENTS TO GRADUALLY INCREASE OVER TIME. DUE TO THE CONDITION OF THE RETURNED LEAD, MANUAL IMPEDANCE TESTING WAS NOT ABLE TO BE PERFORMED; HOWEVER, PREVIOUS LABORATORY TESTING ON LEADS WITH SIMILAR CALCIFICATION FOUND THAT AS THE CALCIFIED MATERIAL IS REMOVED, THE IMPEDANCE MEASUREMENTS DECREASE. IT WAS SUSPECTED THE CALCIFIED MATERIAL ON THE HELIX HAD CAUSED THE INCREASE IN PACING THRESHOLD MEASUREMENTS THAT HAD BEEN OBSERVED DURING THE DEVICE REPLACEMENT PROCEDURE.
(B)(4). THE DEVICE AND LEAD REMAIN IN SERVICE, PENDING THE REPLACEMENT PROCEDURE. THIS REPORT WILL BE UPDATED WHEN ADDITIONAL INFORMATION IS RECEIVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE DEVICE FOLLOW-UP, A REVIEW OF THE LEAD DAILY MEASUREMENTS SHOWED THIS RIGHT VENTRICULAR (RV) LEAD HAD TWO OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS THAT WERE GREATER THAN 200 OHMS. NO OTHER ISSUES WERE NOTED, AND NO CHANGES WERE MADE TO THE SYSTEM AT THIS TIME. LATER, THE PATIENT WAS SEEN AGAIN SO ADDITIONAL SHOCK IMPEDANCE TESTING COULD BE PERFORMED. ALSO, A COPY OF THE DEVICE DATA WAS SAVED TO A DISK AND SUBMITTED TO TECHNICAL SERVICES FOR EVALUATION. UPON ENGINEERING REVIEW OF THE DEVICE DATA, IT WAS CONFIRMED THAT THE SHOCK IMPEDANCE MEASUREMENTS WERE HIGH, OUT-OF-RANGE IN ALL SHOCK CONFIGURATIONS. NO NOISE WAS OBSERVED, AND IT WAS REPORTED THAT AN X-RAY SHOWED NO VISIBLE LEAD DAMAGE. TECHNICAL SERVICES DISCUSSED THE DATA POINTED TO A POTENTIAL LEAD PROBLEM, BUT THERE WAS NOT ENOUGH EVIDENCE TO RULE OUT A DEVICE ISSUE; THEREFORE, IT WAS RECOMMENDED TO REPLACE BOTH THE LEAD AND THE DEVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
--
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 383556 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | LWS | CPI - DEL CARIBE | 0185 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | T167| E143| 0185| 4096 |