TELIGEN
Report
- Report Number
- 2124215-2014-11358
- Event Type
- Injury
- Date Received
- July 1, 2014
- Date of Event
- April 26, 2014
- Report Date
- May 29, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Removal / Correction Number
- Z-0088-2015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
(B)(4). ALL AVAILABLE INFORMATION INDICATES THAT THIS PRODUCT REMAINS IN SERVICE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS NOT INCLUDED IN THE AUGUST 29, 2013 ADVISORY POPULATION, BUT WAS ADDED TO THE EXPANDED POPULATION ON SEPTEMBER 17, 2014.
(B)(4). ADDITIONAL INFORMATION WAS RECEIVED INDICATING THIS DEVICE WAS EXPLANTED. UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. REVIEW OF THE DEVICE MEMORY CONFIRMED THREE LOW VOLTAGE ALERTS (CODE 1003) HAD BEEN RECORDED. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. INITIAL AUTOMATED TESTING VERIFIED BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH COMPROMISED LOW VOLTAGE (BYPASS) CAPACITORS CONNECTED TO THE DEVICE¿S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH-VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL. ON (B)(4), 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS NOT A PART OF THE IDENTIFIED POPULATION.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DISPLAYED THE FAULT CODE 1003 AND THE WARNING MESSAGE THAT THE VOLTAGE WAS TOO LOW FOR THE PROJECTED REMAINING CAPACITY. A MEMORY DOWNLOAD WAS PERFORMED AND SENT TO BOSTON SCIENTIFIC FOR FURTHER EVALUATION. A BOSTON SCIENTIFIC ENGINEER REVIEWED THE DATA AND CONFIRMED THAT A LOW VOLTAGE FAULT HAD BEEN DECLARED AS A RESULT OF AN ADDITIONAL CURRENT DRAIN ON THE DEVICE¿S BATTERY. AT THIS TIME DELIVERY IS CURRENTLY UNAFFECTED; HOWEVER, THE DEVICE IS MALFUNCTIONING AND SHOULD BE REPLACED. THE ICD HAS A SUFFICIENT RESERVE TO MAINTAIN NORMAL THERAPY FUNCTIONS FOR FOUR WEEKS. BEYOND THAT TIME, THERAPY COULD NOT BE GUARANTEED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 384586 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWP | GUIDANT CRM CLONMEL IRELAND | E110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization| L| R | 4087| 1861| 0157| E110| T167 |