FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3905673 · Received May 28, 2014

Report

Report Number
8020893-2014-01283
Event Type
Malfunction
Date Received
May 28, 2014
Date of Event
January 1, 2014
Report Date
April 29, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED ISSUE. THE CSE REPLACED THE MOTHER BOARD AND INSPIRATORY BLIND MATE CABLE TO RESOLVE THE ISSUE. THE DEVICE PASSED ALL CALIBRATIONS, SHORT SELF-TEST (SST), EXTENDED SELF-TEST (EST) AND PERFORMANCE VERIFICATION TESTING ACCORDING TO MFR SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED A DEVICE ALERT WHICH RENDERED THE VENTILATOR INOPERABLE. THERE WAS NO PT INVOLVEMENT AT THE TIME OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
315107 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND 840

Patients

Seq Age Sex Outcome Treatment
1