FDA Adverse Event Injury Summary report: N

INGENIO

MDR report key: 3905651 · Received July 1, 2014

Report

Report Number
2124215-2014-11531
Event Type
Injury
Date Received
July 1, 2014
Date of Event
May 8, 2014
Report Date
July 8, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE RIGHT VENTRICULAR PACING, AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. LABORATORY ANALYSIS CONCLUDED THAT THE ERROR CODES IN DEVICE MEMORY AS WELL AS THE SAFETY MODE OPERATION WERE INDUCED AS A RESULT OF CONTINUOUS APPLICATION OF A MAGNET TO THE DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER EXHIBITED THREE UNSPECIFIED ERROR CODES AND WAS OBSERVED TO BE IN SAFETY MODE UPON INTERROGATION WITH A PROGRAMMER. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) RECOMMENDED DEVICE REPLACEMENT. THE DEVICE WAS EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
383999 INGENIO IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND K173

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| L| R 4285| 1297| 4269| 294-09E| K173