FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 3905590 · Received May 28, 2014

Report

Report Number
1720753-2014-04533
Event Type
Malfunction
Date Received
May 28, 2014
Date of Event
May 7, 2014
Report Date
May 28, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE POWER DISTRIBUTION PCB AND INTERCONNECT CABLE WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXPERIENCED SINGLE BOARD COMPUTER LOCK UPS. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
315054 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1