FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3905587
·
Received May 28, 2014
Report
- Report Number
- 1720753-2014-04523
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- May 6, 2014
- Report Date
- May 28, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SYSTEM BATTERIES WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A PRECHARGE VOLTAGE ERROR MESSAGE. THIS ERROR WILL PREVENT THE SYSTEM FROM BOOTING, RESULTING IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 315053 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |