FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3905580
·
Received May 28, 2014
Report
- Report Number
- 1720753-2014-04507
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- May 6, 2014
- Report Date
- May 28, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FUSE WAS RESEATED AND THE CONNECTORS WERE CLEANED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE FE REPORTED THAT THE SYSTEM INTERMITTENTLY LOCKED UP. THIS CAUSED AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THIS COULD RESULT IN A PROCEDURAL DELAY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314891 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |