ADVANIX? BILIARY
Report
- Report Number
- 3005099803-2014-02349
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- April 15, 2014
- Report Date
- June 11, 2014
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- FGE
- PMA / PMN Number
- K101786
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
PATIENT REPORTED TO BE OVER 18 YEARS OF AGE. EVENT REVEALED BY THE INVESTIGATION OF STENT TORN. THE INVESTIGATION FOUND OUT THAT THE STENT HAD SMALL TEARS ON THE PROXIMAL END AND WAS BENT/KINKED IN SEVERAL PLACES THROUGHOUT. THE NOTED DAMAGE IS LIKELY DUE TO PROCEDURAL/ANATOMICAL FACTORS ENCOUNTERED DURING THE PROCEDURE WHICH MAY HAVE LIMITED THE OVERALL PERFORMANCE OF THE DEVICE, THEREFORE THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY DOUBLE PIGTAIL STENT WAS USED DURING A STENT PLACEMENT PROCEDURE PERFORMED IN THE COMMON BILE DUCT ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THERE WAS STRONG RESISTANCE WHEN THE PHYSICIAN TRIED TO RELEASE THE STENT. THE PHYSICIAN PULLED THE DEVICE WITH STRONG FORCE, THUS, SEPARATING THE INNER CATHETER. THE BROKEN INNER CATHETER DID NOT REMAIN IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER ADVANIX BILIARY DOUBLE PIGTAIL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. INVESTIGATION RESULTS REVEALED THE STENT TO BE TORN, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 385022 | ADVANIX? BILIARY | CATHETER, BILIARY, DIAGNOSTIC | FGE | BOSTON SCIENTIFIC - SPENCER | M00532190 | 16769486 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |