CADIERE FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04002
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- June 13, 2014
- Report Date
- June 25, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THE INSTRUMENT'S GRIP TIPS TO BE BENT. ONE GRIPS WAS BENT WHICH CAUSED SIDE TO SIDE MISALIGNMENT OF THE GRIP. THERE WAS A 0.148 OFFSET AT THE TIPS, INDICATING OVERLOADING AT THE TIP. THE INSTRUMENT PASSED THE ELECTRICAL CONTINUITY TEST. FAILURE ANALYSIS CONCLUDED THAT THE BENT DAMAGE MAY BE DUE TO MISHANDLING/MISUSE. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE FOLLOWING DAMAGES: THE INSTRUMENT'S DISTAL PULLEY HAD MECHANICAL INDENTATIONS/BURRS. THERE WERE INDENTATIONS AT THE EDGE OF THE DISTAL PULLEY AND VISIBLE SCRATCHES ON THE SURFACE OF THE PULLEY. FAILURE ANALYSIS CONCLUDED THAT THE MECHANICAL INDENTATIONS/BURRS DAMAGE MAY BE DUE TO MISHANDLING/MISUSE. IN ADDITION, THE INSTRUMENT'S MAIN TUBE HAD SCRATCH MARKS/ABRASIONS. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WHICH EXHIBITED LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. THE SCRATCHES WERE SHORT IN LENGTH AND WAS NOT AXIALLY WITH THE TUBE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE VARIOUS SCRATCH MARKS ON THE MAIN TUBE WITH LIGHT MATERIAL REMOVAL FOUND DURING FAILURE ANALYSIS IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE TIPS ON THE CADIERE FORCEPS INSTRUMENT WERE CROSSED/MISALIGNED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 384133 | CADIERE FORCEPS INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420049-08 | M10140131 056 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |