FDA Adverse Event Injury Summary report: N

THINLINE II

MDR report key: 3905344 · Received July 1, 2014

Report

Report Number
2124215-2014-11753
Event Type
Injury
Date Received
July 1, 2014
Date of Event
December 2, 2013
Report Date
May 6, 2014
Manufacturer
CPI - DEL CARIBE
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ACCORDING TO AVAILABLE INFORMATION, THIS LEAD WAS SURGICALLY ABANDONED AND REPLACED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A NON-CARDIAC RELATED HOSPITALIZATION, THIS RIGHT VENTRICULAR LEAD WAS OBSERVED TO HAVE LOSS OF CAPTURE. FURTHER INTERROGATION WAS PERFORMED. ALTHOUGH CONDUCTION OF ATRIAL DEFLECTION TO THE VENTRICLE WAS OBSERVED, INTRINSIC QRS OF 50BPM WAS OBSERVED AND A WENCKEBACH SECOND-DEGREE ATRIOVENTRICULAR BLOCK WAS OBSERVED WITH ATRIAL PACING OF 70PPM. THE PATIENT WAS ASYMPTOMATIC. MEASUREMENTS IN BIPOLAR REVEALED A HIGH OUT OF RANGE IMPEDANCE MEASUREMENT. IN ADDITION, THE LEAD WAS NOT PACING OR SENSING APPROPRIATELY. UNIPOLAR MEASUREMENTS WERE NOT OBTAINED. A REVIEW OF THE DATA DETERMINED HIGH MEASUREMENTS APPROXIMATELY FIVE MONTHS EARLIER AND BEEN VARIABLE SINCE THEN. AN X-RAY DID NOT REVEAL ANY ISSUES. THE DEVICE WAS PROGRAMMED TO AAI/50 PPM. THE FOLLOWING DAY, A LEAD REVISION WAS PERFORMED. PRIOR TO LOOSENING THE SET SCREWS, IT WAS VERIFIED THE CONNECTION HAD BEEN SECURED. MEASUREMENTS WITH THE PACING SYSTEM ANALYZER CONFIRMED THE OUT OF RANGE IMPEDANCE MEASUREMENTS. A LEAD FRACTURE WAS SUSPECTED. THIS LEAD WAS SURGICALLY ABANDONED AND REPLACED. POST PROCEDURE, ACCEPTABLE MEASUREMENTS WERE OBTAINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
382947 THINLINE II IMPLANTABLE LEAD LWS CPI - DEL CARIBE 430-35S-52

Patients

Seq Age Sex Outcome Treatment
1 86 YR Hospitalization| L| R MISMATCH| (B)(4)| (B)(4)