Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE REMAINS IN SERVICE. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH THE DEVICE¿S ABILITY TO DELIVER THERAPY WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. REVIEW OF THE DEVICE MEMORY INDICATED THAT A VOLTAGE ALERT WAS RECORDED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE¿S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL. ON AUGUST 29, 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS A PART OF THE IDENTIFIED POPULATION.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SUBSEQUENTLY EXPLANTED AND REPLACED BASED ON BOSTON SCIENTIFIC'S RECOMMENDATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE HAS NOT BEEN RETURNED. SHOULD THE DEVICE GET RETURNED OR SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. THE PATIENT REPORTED DIFFICULTY WITH HEARING, SO WAS NOT AWARE IF THERE WERE BEEP TONES FROM THE DEVICE. DISK ANALYSIS SHOWED THE BATTERY APPEARS TO BE DEPLETING MORE QUICKLY THAN EXPECTED. DEVICE REPLACEMENT WAS RECOMMENDED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
383227 COGNIS IMPLANTABLE CHF PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND N118

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization| L| R N118| 1688TC| 0186| 4518