FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 3905224
·
Received May 28, 2014
Report
- Report Number
- 2518422-2014-00870
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- May 6, 2014
- Report Date
- May 6, 2014
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- CBK
- PMA / PMN Number
- K093416
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S POWER MANAGEMENT BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314818 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC. | 1032800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |