FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3905201 · Received May 28, 2014

Report

Report Number
2518422-2014-00874
Event Type
Malfunction
Date Received
May 28, 2014
Date of Event
May 7, 2014
Report Date
May 7, 2014
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A FAILURE OF THE DEVICE TO POWER ON WITH AC POWER WAS OBSERVED. THE DEVICE'S POWER SUPPLY WAS REPLACED TO ADDRESS THE ISSUE. THE MALFUNCTION OF THE POWER SUPPLY WOULD RESULT IN THE FAILURE OF THE DEVICE TO OPERATE ON AC POWER OR TO CHARGE ITS INTERNAL BATTERY. THIS PARTICULAR MALFUNCTION FOR THE POWER SUPPLY HAS ONLY BEEN REPORTED ON DEVICES THAT ARE BEING SOLD INTERNATIONALLY FOR USE WITH 220 VOLT POWER SOURCES.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314910 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054096

Patients

Seq Age Sex Outcome Treatment
1