FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 3905179 · Received June 10, 2014

Report

Report Number
1824206-2014-01749
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BED WAS NOT THE ISSUE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN TESTED THE BED IN A DIFFERENT ROOM AND CONFIRMED THE NON-HILL-ROM NURSE CALL SYSTEM WAS THE ISSUE IN THAT ROOM. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED WOULD NOT SEND A CALL TO THE NURSES' STATION. THE BED WAS LOCATED IN CLC ROOM 60 OF THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339613 VERSACARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200