FDA Adverse Event Malfunction Summary report: N

ETS FLEX ARTICNG LNR CUTR

MDR report key: 3905162 · Received July 1, 2014

Report

Report Number
3005075853-2014-04532
Event Type
Malfunction
Date Received
July 1, 2014
Date of Event
May 15, 2014
Report Date
June 6, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE REPORTED EVENT COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DIFFICULTIES NOTED DURING THE FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTAT. INFORMATION IS UNAVAILABLE; DEVICE WAS NOT RETURNED FOR EVALUATION. (B)(4).

Description of Event or Problem · 1

PER MEDWATCH (B)(4) ATTACHED TO THIS REPORT. IT WAS REPORTED THAT DURING AN APPENDECTOMY PROCEDURE, THERE WAS A CARTRIDGE LOCKOUT. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
383514 ETS FLEX ARTICNG LNR CUTR STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4EJ22

Patients

Seq Age Sex Outcome Treatment
1