FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 3904881 · Received June 4, 2014

Report

Report Number
3008642652-2014-01679
Event Type
Malfunction
Date Received
June 4, 2014
Date of Event
April 28, 2014
Report Date
May 30, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CONNECTOR) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE TRUNK CABLE CONNECTOR WAS DAMAGED, BREAKING THE CONNECTION BETWEEN THE CONNECTOR WIRES AND CONNECTOR PINS. THE ROOT CAUSE FOR THE DAMAGE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER ELECTRODE BELT CONNECTOR WAS DAMAGED. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
327900 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR