FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3904854 · Received June 4, 2014

Report

Report Number
3008642652-2014-01676
Event Type
Malfunction
Date Received
June 4, 2014
Date of Event
April 26, 2014
Report Date
May 30, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (NOT CHARGING BATTERIES) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE BATTERY CHARGER WAS UNABLE TO DETECT INSERTED BATTERY PACKS. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A SHORTED U13 CMOS FLASH MICROCONTROLLER. THE ROOT CAUSE FOR THE SHORTED U13 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS BATTERY CHARGER WASN'T CHARGING THIS BATTERIES. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
327880 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 52 YR