FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3904854
·
Received June 4, 2014
Report
- Report Number
- 3008642652-2014-01676
- Event Type
- Malfunction
- Date Received
- June 4, 2014
- Date of Event
- April 26, 2014
- Report Date
- May 30, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (NOT CHARGING BATTERIES) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE BATTERY CHARGER WAS UNABLE TO DETECT INSERTED BATTERY PACKS. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A SHORTED U13 CMOS FLASH MICROCONTROLLER. THE ROOT CAUSE FOR THE SHORTED U13 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS BATTERY CHARGER WASN'T CHARGING THIS BATTERIES. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 327880 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |