COGNIS
Report
- Report Number
- 2124215-2014-10026
- Event Type
- Injury
- Date Received
- July 1, 2014
- Date of Event
- April 21, 2014
- Report Date
- July 15, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Removal / Correction Number
- Z-0087-2015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS CURRENTLY AVAILABLE, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS NOT INCLUDED IN THE AUGUST 29, 2013 ADVISORY POPULATION, BUT WAS ADDED TO THE EXPANDED POPULATION ON SEPTEMBER 17, 2014.
(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH THE DEVICE¿S ABILITY TO DELIVER THERAPY WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. REVIEW OF THE DEVICE MEMORY INDICATED THAT NO VOLTAGE ALERT WAS RECORDED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE¿S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL. ON AUGUST 29, 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS NOT A PART OF THE IDENTIFIED POPULATION.
(B)(4).
--
--
ADDITIONAL INFORMATION WAS RECEIVED THAT THIS DEVICE WAS SUCCESSFULLY EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS.
AS PART OF AN ANALYSIS TO REVIEW THE SAFETY ARCHITECTURE LOW VOLTAGE ALERT (CODE 1003), BATTERY VOLTAGE DATA WAS PULLED FROM THE REMOTE PATIENT MONITORING SYSTEM. DURING OUR ANALYSIS, THIS DEVICE WAS IDENTIFIED AS DEMONSTRATING EVIDENCE OF PREMATURE BATTERY DEPLETION THAT COULD POTENTIALLY LEAD TO COMPROMISED THERAPY IF NOT ADDRESSED IN A TIMELY MANNER. BOSTON SCIENTIFIC TECHNICAL SERVICES PROACTIVELY REACHED OUT TO THE FIELD REPRESENTATIVE TO NOTIFY THEM OF THIS FINDING AND RECOMMEND DEVICE REPLACEMENT. BOSTON SCIENTIFIC THEN RECEIVED ADDITIONAL INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. DEVICE REPLACEMENT WAS SCHEDULED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. AT THIS TIME, AVAILABLE INFORMATION SUGGESTS THE DEVICE REMAINS IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 383084 | COGNIS | IMPLANTABLE CHF PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | N119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Hospitalization| L| R | 4470| N119| 4542| 0185 |