INCEPTA
Report
- Report Number
- 2124215-2014-11083
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- April 28, 2014
- Report Date
- April 28, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EMITTED BEEPING TONES. UPON THE PATIENT'S PRESENTATION FOR FOLLOW-UP IT WAS DISCOVERED THAT A SINGLE HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT GREATER THAN 125 OHMS WAS RECORDED TO THE DEVICE. UPON FURTHER REVIEW, SHOCK IMPEDANCE MEASUREMENTS APPEARED TO BE TRENDING HIGHER OVER PRECEDING MONTHS. ALL OTHER LEAD PARAMETERS WERE NOTED TO BE STABLE AND WITHIN NORMAL LIMITS. POCKET MANIPULATIONS WERE PERFORMED WITH THE PATIENT IN DIFFERENT POSITIONS AND SHOCK IMPEDANCES WERE NOTED TO VARY 10-20 OHMS WHICH THE PHYSICIAN CONSIDERED NORMAL. NO COMMANDED SHOCKS WERE PERFORMED AND THERE WAS NO SUSPICION OF PHYSICAL LEAD DAMAGE. THE PATIENT WILL BE SEEN FOR FOLLOW UP IN THREE MONTHS TIME. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS SYSTEM REMAINS IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 383083 | INCEPTA | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | F162 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | F162| 0292| 4470 |