TELIGEN
Report
- Report Number
- 2124215-2014-10419
- Event Type
- Injury
- Date Received
- July 1, 2014
- Date of Event
- April 22, 2014
- Report Date
- May 16, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Removal / Correction Number
- Z-0088-2015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. REVIEW OF THE DEVICE MEMORY CONFIRMED THAT A LOW VOLTAGE ALERT (CODE 1003) HAD BEEN RECORDED. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. THE DEVICE WAS THEN SUBJECTED TO AUTOMATED TESTING TO VERIFY PROPER OPERATION. THE DEVICE WOULD NOT PASS DUE TO A LOW BATTERY. AFTER THE TITANIUM CASE WAS OPENED, MICROSCOPIC VISUAL INSPECTION DID NOT REVEAL ANY IRREGULARITIES. THE BATTERY WAS THEN REMOVED, SO THAT AN EXTERNAL POWER SUPPLY COULD BE CONNECTED TO THE DEVICE FOR FURTHER ANALYSIS. A HIGHER THAN NORMAL CURRENT USAGE WAS OBSERVED. ELECTRICAL TESTING ISOLATED THE HIGH CURRENT CONDITION TO TWO COMPROMISED LOW VOLTAGE CAPACITORS THAT ARE CONNECTED TO THE DEVICE¿S BATTERY. THIS TYPE OF COMPONENT MALFUNCTION CAN RESULT IN A HIGH CURRENT DRAIN, WHICH OVER TIME CAN RESULT IN PREMATURE BATTERY DEPLETION, AS WAS OBSERVED IN THIS CASE. ON (B)(4) 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS NOT A PART OF THE IDENTIFIED POPULATION.
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS NOT INCLUDED IN THE AUGUST 29, 2013 ADVISORY POPULATION, BUT WAS ADDED TO THE EXPANDED POPULATION ON SEPTEMBER 17, 2014.
(B)(4). THE DEVICE IS CURRENTLY UNDERGOING LABORATORY TESTING TO DETERMINE ROOT CAUSE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LOCAL REPRESENTATIVE CONTACTED BOSTON SCIENTIFIC'S TECHNICAL SERVICES (TS). ACCORDING TO THE CLINIC NURSE, THE PATIENT HAD BEEN ADMITTED INTO THE HOSPITAL DUE TO CHEST PAIN (NON-DEVICE RELATED ISSUE). THE DEVICE WAS INTERROGATED AND A CODE#1003 WAS DISPLAYED. DATA FROM DEVICE WAS RECEIVED AT BOSTON SCIENTIFIC FOR ANALYSIS. IT WAS CONFIRMED THROUGH DATA ANALYSIS THAT A LOW VOLTAGE FAULT HAD BEEN DECLARED IN (B)(6) 2014. THE VOLTAGE IS CURRENTLY 3.019 VOLTS AND THERAPY DELIVERY IS UNAFFECTED. THE DEVICE HARDWARE IS NOT DETECTING THE LOSS OF BATTERY ENERGY. BECAUSE OF THIS, THE BATTERY STATUS INDICATORS ARE NOT REFLECTING THE DEPLETION CONDITION AND ARE INACCURATE. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, THE DAILY DEVICE POWER FLUCTUATIONS WERE ESTIMATED. TO DATE, THE CURRENT APPEARS STEADY; HOWEVER, THIS BEHAVIOR MAY CHANGE UNPREDICTABLY. AT THIS POINT, THE BATTERY DOES HAVE A SIGNIFICANT RESERVE CAPACITY. THE DATA INDICATES THAT THERE IS SUFFICIENT RESERVE FOR THE DEVICE TO MAINTAIN NORMAL THERAPY FUNCTIONS FOR 28 DAYS' TIME. BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT WAS PRESENTED BACK TO THE ELECTROPHYSIOLOGY (EP) LABORATORY. THE DEVICE WAS EXPLANTED AND REPLACED WITHOUT INCIDENT.
--
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 382295 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWP | GUIDANT CRM CLONMEL IRELAND | E102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Hospitalization| L| R | 0185| E102 |