COGNIS
Report
- Report Number
- 2124215-2014-10403
- Event Type
- Injury
- Date Received
- July 1, 2014
- Date of Event
- March 28, 2014
- Report Date
- May 13, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
(B)(4). UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. REVIEW OF THE DEVICE MEMORY CONFIRMED LOW VOLTAGE ALERTS (CODE 1003) HAD BEEN RECORDED. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED NO ANOMALIES. PACING PULSES WERE PRESENT, BUT FULL PRODUCT RETURN TESTING WAS NOT PERFORMED DUE TO THE LOW BATTERY VOLTAGE. AFTER THE TITANIUM CASE WAS OPENED, MICROSCOPIC VISUAL INSPECTION DID NOT REVEAL ANY IRREGULARITIES. THE BATTERY WAS THEN REMOVED, SO THAT AN EXTERNAL POWER SUPPLY COULD BE CONNECTED TO THE DEVICE FOR FURTHER ANALYSIS. A HIGHER THAN NORMAL CURRENT USAGE WAS OBSERVED. ELECTRICAL TESTING ISOLATED THE HIGH CURRENT CONDITION TO TWO COMPROMISED LOW VOLTAGE CAPACITORS THAT ARE CONNECTED TO THE DEVICE¿S BATTERY. THIS TYPE OF COMPONENT MALFUNCTION CAN RESULT IN A HIGH CURRENT DRAIN, WHICH OVER TIME CAN RESULT IN PREMATURE BATTERY DEPLETION, AS WAS OBSERVED IN THIS CASE. ON (B)(4) 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS NOT A PART OF THE IDENTIFIED POPULATION.
(B)(4). THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. DEVICE REPLACEMENT WAS RECOMMENDED. THE DEVICE WAS EXPLANTED AND REPLACED THE SAME DAY. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 382226 | COGNIS | IMPLANTABLE CHF PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | N119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Hospitalization| L| R | 4591| 4592| N141| 4538| 4076| 0185| N119 |