TELIGEN
Report
- Report Number
- 2124215-2014-11672
- Event Type
- Injury
- Date Received
- July 1, 2014
- Date of Event
- April 9, 2014
- Report Date
- July 11, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). ADDITIONAL ANALYSIS HAS BEEN COMPLETED AND NO CURRENT LEAKAGE IN THE BATTERY ITSELF WAS IDENTIFIED. DURING TESTING NO CHANGES IN THE BATTERY CURRENT DRAW WERE NOTED. DETAILED ANALYSIS OF INTERNAL COMPONENTS WAS UNABLE TO IDENTIFY A ROOT CAUSE FOR THE CONFIRMED PREMATURE BATTERY DEPLETION.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS ICD WAS THOROUGHLY INSPECTED AND ANALYZED. EXTERNAL VISUAL INSPECTION OF THE DEVICE REVEALED NO ANOMALIES. REVIEW OF THE DEVICE MEMORY CONFIRMED THAT A LOW VOLTAGE ALERT WAS RECORDED. ALTHOUGH THE DEVICE MEMORY DIAGNOSTICS DEMONSTRATED THAT THE DAILY BATTERY VOLTAGE MEASUREMENTS DISPLAYED AN IRREGULAR PATTERN OF DISCHARGE, THE BATTERY VOLTAGE LEVEL AT EXPLANT 2.988 VOLTS WAS SUFFICIENT TO ENSURE THERAPY AVAILABILITY/DELIVERY WHILE THE DEVICE WAS IMPLANTED. THE DEVICE CASE WAS THEN OPENED TO FACILITATE ANALYSIS OF THE INTERNAL COMPONENTS. THE BATTERY WAS SEPARATED FROM THE OTHER DEVICE ELECTRONICS AND THE OVERALL CURRENT DRAW OF THE CIRCUITRY WAS MEASURED. A NORMAL CURRENT DRAIN WAS OBSERVED WITHIN THE CIRCUITRY. COLLECTIVELY, THE PATTERN OF IRREGULAR DAILY BATTERY VOLTAGE MEASUREMENTS IN CONJUNCTION WITH NORMAL POWER LEVELS AND DEVICE HYBRID CURRENT DRAW IS CONSISTENT WITH BEHAVIOR OF DEVICES WHERE A LATENT CURRENT LEAKAGE PATH HAS OCCURRED WITHIN THE BATTERY ITSELF, RESULTING IN A PARTIAL DEPLETION OF THE BATTERY.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. SUBSEQUENTLY THE DEVICE WAS EXPLANTED FOR PREMATURE BATTERY DEPLETION AND RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 382225 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWP | GUIDANT CRM CLONMEL IRELAND | E110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Hospitalization| L| R | E110| 1600| 1745| 4086| 1857| 1550| 0184| 4312| 0041| T180 |