FDA Adverse Event Injury Summary report: N

TELIGEN

MDR report key: 3904745 · Received July 1, 2014

Report

Report Number
2124215-2014-11672
Event Type
Injury
Date Received
July 1, 2014
Date of Event
April 9, 2014
Report Date
July 11, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL ANALYSIS HAS BEEN COMPLETED AND NO CURRENT LEAKAGE IN THE BATTERY ITSELF WAS IDENTIFIED. DURING TESTING NO CHANGES IN THE BATTERY CURRENT DRAW WERE NOTED. DETAILED ANALYSIS OF INTERNAL COMPONENTS WAS UNABLE TO IDENTIFY A ROOT CAUSE FOR THE CONFIRMED PREMATURE BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS ICD WAS THOROUGHLY INSPECTED AND ANALYZED. EXTERNAL VISUAL INSPECTION OF THE DEVICE REVEALED NO ANOMALIES. REVIEW OF THE DEVICE MEMORY CONFIRMED THAT A LOW VOLTAGE ALERT WAS RECORDED. ALTHOUGH THE DEVICE MEMORY DIAGNOSTICS DEMONSTRATED THAT THE DAILY BATTERY VOLTAGE MEASUREMENTS DISPLAYED AN IRREGULAR PATTERN OF DISCHARGE, THE BATTERY VOLTAGE LEVEL AT EXPLANT 2.988 VOLTS WAS SUFFICIENT TO ENSURE THERAPY AVAILABILITY/DELIVERY WHILE THE DEVICE WAS IMPLANTED. THE DEVICE CASE WAS THEN OPENED TO FACILITATE ANALYSIS OF THE INTERNAL COMPONENTS. THE BATTERY WAS SEPARATED FROM THE OTHER DEVICE ELECTRONICS AND THE OVERALL CURRENT DRAW OF THE CIRCUITRY WAS MEASURED. A NORMAL CURRENT DRAIN WAS OBSERVED WITHIN THE CIRCUITRY. COLLECTIVELY, THE PATTERN OF IRREGULAR DAILY BATTERY VOLTAGE MEASUREMENTS IN CONJUNCTION WITH NORMAL POWER LEVELS AND DEVICE HYBRID CURRENT DRAW IS CONSISTENT WITH BEHAVIOR OF DEVICES WHERE A LATENT CURRENT LEAKAGE PATH HAS OCCURRED WITHIN THE BATTERY ITSELF, RESULTING IN A PARTIAL DEPLETION OF THE BATTERY.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. SUBSEQUENTLY THE DEVICE WAS EXPLANTED FOR PREMATURE BATTERY DEPLETION AND RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
382225 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| L| R E110| 1600| 1745| 4086| 1857| 1550| 0184| 4312| 0041| T180