THERMOCOOL® SF NAV UNI-DIRECTIONAL CATHETER
Report
- Report Number
- 9673241-2014-00239
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- June 10, 2014
- Report Date
- June 10, 2014
- Manufacturer
- BIOSENSE WEBSTER, INC. (JUAREZ)
- Product Code
- LPB
- PMA / PMN Number
- P030031/S034
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE BWI FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETED AT THIS TIME. (B)(4).
(B)(4) IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION PROCEDURE WITH A THERMOCOOL SF NAV UNI-DIRECTIONAL CATHETER AND HIGH IMPEDANCE AND CHAR WERE NOTICED ON THE DEVICE. DURING RF ABLATION, CHAR WAS NOTICED TOWARD THE END OF THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SIMILAR-LIKE DEVICE. THE BWI FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION, VISUALLY INSPECTED WAS PERFORMED AND RED MATERIAL SIMILAR TO HUMAN BLOOD WAS FOUND ON THE TIP DOME. NO CHAR WAS OBSERVED. PER THE EVENT, THE CATHETER WAS TESTED FOR ELECTRICAL PERFORMANCE, TEMPERATURE RESPONSE AND STOCKERT COMPATIBILITY AND WAS FOUND WITHIN SPECIFICATIONS. IN ADDITION, AN IRRIGATION TEST WAS PERFORMED AND CATHETER PASSED, NO OCCLUSION WAS OBSERVED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE CUSTOMER COMPLAINT WAS NOT CONFIRMED.
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATING PROCEDURE WITH A THERMOCOOL® SF NAV UNI-DIRECTIONAL CATHETER AND HIGH IMPEDANCE AND CHAR WERE NOTICED ON THE DEVICE. DURING RF ABLATION, CHAR WAS NOTICED TOWARD THE END OF THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SIMILAR-LIKE DEVICE. SETTINGS DURING THE EVENT INCLUDE: IMPEDANCE CUT-OFF AT 250 OHMS, IN POWER CONTROL MODE AT 30 WATTS AND 8 CC/MIN OF FLOW RATE. THIS EVENT IS BEING REPORTED DUE TO THE POTENTIAL RISK OF THE CHAR FOUND TO THE PATIENT. THE AWARENESS DATE WAS UPDATED FOR THIS RECORD BECAUSE MORE INFORMATION BECAME AVAILABLE TO BWI ON JUNE 12TH, 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 383201 | THERMOCOOL® SF NAV UNI-DIRECTIONAL CATHETER | CARDIAC ABLATION PERCUTANEOUS CATHETER | LPB | BIOSENSE WEBSTER, INC. (JUAREZ) | D-1315-04-S | 16064837L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |