FDA Adverse Event Malfunction Summary report: N

POSEY SITTER ELITE

MDR report key: 3904678 · Received May 22, 2014

Report

Report Number
2020362-2014-00192
Event Type
Malfunction
Date Received
May 22, 2014
Report Date
April 3, 2014
Manufacturer
J. T. POSEY COMPANY
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: EVALUATION OF THE RETURNED PRODUCT CONFIRMED THE REPORT ISSUED. THE UNIT NURSE CALL RECEPTACLE OUTER HOUSING IS DAMAGED, BUT THE INTERNAL MECHANISM IS STILL INTACT. ALTHOUGH THE NURSE CALL RECEPTACLE IS DAMAGED, UNIT CAN STILL SECURE THE NURSE CALL CABLE. HOWEVER, THE RED LED AT THE NURSE CALL TEST FIXTURE TURNS ON AND OFF WHEN THE NURSE CALL CABLE IS MOVED. ALSO, A RATTLING SOUND IS HEARD COMING FROM INSIDE OF THE UNIT. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THE NURSE CALL RECEPTACLE ON THE ALARM IS DAMAGED. THE ISSUE WAS DISCOVERED DURING SETUP, BUT THE DATE IS UNKNOWN. NO PATIENT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
306890 POSEY SITTER ELITE KMI J. T. POSEY COMPANY 8345 NA

Patients

Seq Age Sex Outcome Treatment
1 NA NURSE CALL CABLE, MODEL 8282, LOT# UNKNOWN