FDA Adverse Event Injury Summary report: N

COGNIS

MDR report key: 3904570 · Received July 1, 2014

Report

Report Number
2124215-2014-09807
Event Type
Injury
Date Received
July 1, 2014
Date of Event
April 15, 2014
Report Date
May 15, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Removal / Correction Number
Z-0025-2014
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RETURN, THE DEVICE WILL UNDERGO DETAILED ANALYSIS TO DETERMINE ROOT CAUSE.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. REVIEW OF THE DEVICE MEMORY CONFIRMED THAT NO LOW VOLTAGE ALERTS (CODE 1003) HAD BEEN RECORDED. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED TOOL MARKS ON THE CASE AND BODILY FLUID CONTAMINATION IN THE LEAD BARRELS. THE HEADER TO CASE BOND AND THE SEAL PLUGS WERE INTACT. ALL OF THE SETSCREWS OPERATED NORMALLY. INITIAL AUTOMATED TESTING VERIFIED BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH COMPROMISED LOW VOLTAGE (BYPASS) CAPACITORS CONNECTED TO THE DEVICE¿S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH-VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL. ON (B)(6) 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS A PART OF THE IDENTIFIED POPULATION.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS RECEIVED AT BOSTON SCIENTIFIC'S RETURN PRODUCTS DEPARTMENT.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

AS PART OF AN ANALYSIS TO REVIEW THE SAFETY ARCHITECTURE LOW VOLTAGE ALERT (CODE 1003), BATTERY VOLTAGE DATA WAS PULLED FROM THE REMOTE PATIENT MONITORING SYSTEM. DURING OUR ANALYSIS, THE DEVICE WAS IDENTIFIED AS DEMONSTRATING EVIDENCE OF PREMATURE BATTERY DEPLETION THAT COULD POTENTIALLY LEAD TO COMPROMISED THERAPY IF NOT ADDRESSED IN A TIMELY MANNER. BOSTON SCIENTIFIC TECHNICAL SERVICES PROACTIVELY REACHED OUT TO THE "FIELD REPRESENTATIVE/PHYSICIAN" TO NOTIFY THEM OF THIS FINDING AND RECOMMEND DEVICE REPLACEMENT. AT THIS TIME, THE DEVICE REMAINS IMPLANTED AND NO ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT WAS PRESENTED BACK TO THE ELECTROPHYSIOLOGY (EP) LABORATORY. THE DEVICE WAS EXPLANTED WITHOUT INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
382352 COGNIS IMPLANTABLE CHF PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 85 YR Hospitalization| L| R 4542| N119| 4035| 1290| 0157| 4064| 4039| 1296