TELIGEN
Report
- Report Number
- 2124215-2014-09636
- Event Type
- Injury
- Date Received
- July 1, 2014
- Date of Event
- April 17, 2014
- Report Date
- July 9, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Removal / Correction Number
- Z-0088-2015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED TOOL MARKS ON THE DEVICE CASE AND HEADER. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH THE DEVICE¿S ABILITY TO DELIVER THERAPY WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. REVIEW OF THE DEVICE MEMORY INDICATED THAT A VOLTAGE ALERT WAS RECORDED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE¿S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE¿S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE¿S BATTERY FASTER THAN NORMAL. ON AUGUST 29, 2013, BOSTON SCIENTIFIC DISTRIBUTED A LETTER TO PHYSICIANS CONCERNING THIS ISSUE. THIS LETTER INFORMED PHYSICIANS THAT, IN A SUBSET OF DEVICES, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS DEVICE IS NOT A PART OF THE IDENTIFIED POPULATION.
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS NOT INCLUDED IN THE AUGUST 29, 2013 ADVISORY POPULATION, BUT WAS ADDED TO THE EXPANDED POPULATION ON SEPTEMBER 17, 2014.
(B)(6). THE DEVICE IS EXPECTED TO BE RETURNED FOR LABORATORY ANALYSIS. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. THIS DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 384944 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWP | GUIDANT CRM CLONMEL IRELAND | E102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| L| R | E102| 0157 |