FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 3904463 · Received May 28, 2014

Report

Report Number
1828100-2014-00462
Event Type
Malfunction
Date Received
May 28, 2014
Date of Event
May 4, 2014
Report Date
May 5, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

USER FACILITY'S BIOMEDICAL ENGINEER (BIOMED) WAS WORKING WITH TECHNICAL SUPPORT TO GET PART NUMBERS FOR THE REPAIR. BIOMED TOOK THE PUMP APART AND IT LOOKS LIKE THE IMPELLERS WERE WOBBLY. THE BIOMED REPLACED THE PUMP AT THE HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT WAS NOISY. THE UNIT WAS INTERMITTENTLY FUNCTIONING. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, OR NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
315004 DUAL COOLER/HEATER SARNS DUAL COOLER/HEATER DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 11160

Patients

Seq Age Sex Outcome Treatment
1