FDA Adverse Event
Malfunction
Summary report: N
DUAL COOLER/HEATER
MDR report key: 3904463
·
Received May 28, 2014
Report
- Report Number
- 1828100-2014-00462
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- May 4, 2014
- Report Date
- May 5, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- PREAMEND
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
USER FACILITY'S BIOMEDICAL ENGINEER (BIOMED) WAS WORKING WITH TECHNICAL SUPPORT TO GET PART NUMBERS FOR THE REPAIR. BIOMED TOOK THE PUMP APART AND IT LOOKS LIKE THE IMPELLERS WERE WOBBLY. THE BIOMED REPLACED THE PUMP AT THE HOSPITAL.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT WAS NOISY. THE UNIT WAS INTERMITTENTLY FUNCTIONING. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, OR NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 315004 | DUAL COOLER/HEATER | SARNS DUAL COOLER/HEATER | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 11160 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |