FDA Adverse Event Malfunction Summary report: N

RECLAIM DIST REAMER EXTENSION

MDR report key: 3904276 · Received July 1, 2014

Report

Report Number
1818910-2014-22280
Event Type
Malfunction
Date Received
July 1, 2014
Date of Event
June 19, 2014
Report Date
June 20, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED DEVICE CONFIRMS THE REPORTED EVENT OF DISTAL TIP BREAKAGE. A SEARCH OF THE COMPLAINTS DATABASES HAS IDENTIFIED A TREND OF THIS ISSUE ON THE (B)(4), REFERENCE CAPA-(B)(4). EVALUATION BY DEPUY METROLOGY, MATERIALS SCIENCE, PRODUCT DEVELOPMENT CONFIRMED THE ISSUE, BUT FOUND THE PRODUCT WAS MADE TO SPECIFICATIONS. A MEDICAL DEVICE CORRECTION NOTICE WAS DISTRIBUTED ON FEBRUARY 25, 2013 FOR ALL LOTS OF THE 2975-00-500 PRODUCT CODE MANUFACTURED SINCE DECEMBER 2011. THE NOTICE INDICATED THAT DEPUY HAS IDENTIFIED THE POTENTIAL FOR THE RECLAIM REAMER EXTENSION TABS TO BREAK. THE REAMER EXTENSIONS ARE NOT IMMEDIATELY BEING REMOVED FROM THE MARKET AND MAY CONTINUE TO BE USED UNTIL A DESIGN CHANGE IS IMPLEMENTED AND NEW DEVICES ARE AVAILABLE. ECO416771 DESIGN VALIDATION WAS IMPLEMENTED ON (B)(4) 2013. DEPUY PRODUCT DEVELOPMENT IS CURRENTLY IN PROCESS OF REDESIGNING THE INSTRUMENT TO UPDATE THE DESIGN AND BOLSTER ITS DURABILITY. THE ROOT CAUSE IS ATTRIBUTED TO DESIGN. MONITOR THROUGH TREND ANALYSIS PER SEP-419.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

TWO DISTAL REAMER EXTENSION INSTRUMENTS WERE BROKEN DURING SURGERY - DAMAGED AT THE DISTAL AND CONNECTING TO DISTAL REAMER WHERE CONNECTION OCCURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
384159 RECLAIM DIST REAMER EXTENSION HIP INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. 1818910 J0111

Patients

Seq Age Sex Outcome Treatment
1 72 YR