RECLAIM DIST REAMER EXTENSION
Report
- Report Number
- 1818910-2014-22280
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- June 19, 2014
- Report Date
- June 20, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 1818910
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
EXAMINATION OF THE RETURNED DEVICE CONFIRMS THE REPORTED EVENT OF DISTAL TIP BREAKAGE. A SEARCH OF THE COMPLAINTS DATABASES HAS IDENTIFIED A TREND OF THIS ISSUE ON THE (B)(4), REFERENCE CAPA-(B)(4). EVALUATION BY DEPUY METROLOGY, MATERIALS SCIENCE, PRODUCT DEVELOPMENT CONFIRMED THE ISSUE, BUT FOUND THE PRODUCT WAS MADE TO SPECIFICATIONS. A MEDICAL DEVICE CORRECTION NOTICE WAS DISTRIBUTED ON FEBRUARY 25, 2013 FOR ALL LOTS OF THE 2975-00-500 PRODUCT CODE MANUFACTURED SINCE DECEMBER 2011. THE NOTICE INDICATED THAT DEPUY HAS IDENTIFIED THE POTENTIAL FOR THE RECLAIM REAMER EXTENSION TABS TO BREAK. THE REAMER EXTENSIONS ARE NOT IMMEDIATELY BEING REMOVED FROM THE MARKET AND MAY CONTINUE TO BE USED UNTIL A DESIGN CHANGE IS IMPLEMENTED AND NEW DEVICES ARE AVAILABLE. ECO416771 DESIGN VALIDATION WAS IMPLEMENTED ON (B)(4) 2013. DEPUY PRODUCT DEVELOPMENT IS CURRENTLY IN PROCESS OF REDESIGNING THE INSTRUMENT TO UPDATE THE DESIGN AND BOLSTER ITS DURABILITY. THE ROOT CAUSE IS ATTRIBUTED TO DESIGN. MONITOR THROUGH TREND ANALYSIS PER SEP-419.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
TWO DISTAL REAMER EXTENSION INSTRUMENTS WERE BROKEN DURING SURGERY - DAMAGED AT THE DISTAL AND CONNECTING TO DISTAL REAMER WHERE CONNECTION OCCURS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 384159 | RECLAIM DIST REAMER EXTENSION | HIP INSTRUMENT/TRIAL | LXH | DEPUY ORTHOPAEDICS, INC. 1818910 | J0111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |