FDA Adverse Event Malfunction Summary report: N

EASYTRAK 2

MDR report key: 3904265 · Received July 1, 2014

Report

Report Number
2124215-2014-08950
Event Type
Malfunction
Date Received
July 1, 2014
Date of Event
December 30, 2011
Report Date
April 10, 2014
Manufacturer
CPI - DEL CARIBE
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD AND ASSOCIATED CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DISPLAYED PACING IMPEDANCES OF GREATER THAN 2000 OHMS. THE IMPEDANCE TREND WAS REVIEWED AND IT WAS NOTED THAT THE IMPEDANCE HAS BEEN FLUCTUATING OVER THE PAST YEAR. THE PATIENT WAS SEEN IN THE CLINIC. LV THRESHOLD MEASUREMENTS HAD INCREASED TO 4.0V @ 2.0MS, FROM 2.0V @ 1.0MS AT IMPLANT. THE PATIENT HAD ALSO BEEN EXPERIENCING DIAPHRAGMATIC STIMULATION. THE LEAD WAS REPROGRAMMED FROM LV TIP TO LV RING TO LV RING TO RV COIL, RESULTING IN PACING IMPEDANCE MEASUREMENTS WITHIN ACCEPTABLE LIMITS AND A THRESHOLD MEASUREMENT OF 3.0V @ 2.0MS. THE PHYSICIAN ELECTED TO CONTINUE MONITORING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
384106 EASYTRAK 2 IMPLANTABLE LEAD NIK CPI - DEL CARIBE 4518

Patients

Seq Age Sex Outcome Treatment
1 76 YR 4518| N118| 0155| 1853| H179| 4480| 1861